The SPOT-IL positron beam construction and its use for Doppler Broadening measurement of annealed Cu

P. Or, L. Chelladurai, D. Cohen, A. Amrosi, T. Cohen, I. Sabo-Napadensky, E. Gordon,S. Cohen, O. Presler, E. O. Cohen, E. Piasetzky, H. Steinberg,S. May-Tal Beck,G. Ron

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS(2024)

引用 0|浏览6
暂无评分
摘要
The construction and successive first operation of a slow positron beam at The Hebrew University of Israel are reported here. The beam follows a traditional design, using a 22Na source, and a Tungsten moderator. It is designed to allow a combined measurement of Doppler-Broadening (DB), sample conductivity, and sample in situ cooling, with the flexibility to add more detection options in the future. The beam energy varies between 0.03 keV and 30 keV. The detection system consists of two high-purity Germanium detectors, facing each other, allowing coincidence DB measurements. Event readout is done using a state-of-the-art compact desktop system. The beam has been successfully tested by using it to characterize annealed Copper (Cu) samples. Three samples were measured at various energies (0.5 keV-16.5 keV), one before any treatment and two annealed at 250 degrees C and 500 degrees C. The results, fitted using VEPFIT, show consistent behavior with previous measurements.
更多
查看译文
关键词
Slow positron beam,Positron annihilation spectroscopy,Doppler Broadening,Construction,Compact DAQ,Copper