Characterization of a Low-Profile, Flexible, and Acoustically Transparent Receive-Only MRI Coil Array for High Sensitivity MR-Guided Focused Ultrasound

IEEE ACCESS(2022)

引用 0|浏览6
暂无评分
摘要
Magnetic resonance guided focused ultrasound (MRgFUS) is a non-invasive therapeutic modality for neurodegenerative diseases that employs real-time imaging and thermometry monitoring of targeted regions. MRI is used in guidance of ultrasound treatment; however, the MR image quality in current clinical applications is poor when using the vendor built-in body coil. We present an 8-channel, ultra-thin, flexible, and acoustically transparent receive-only head coil design (FUS-Flex) to improve the signal-to-noise ratio (SNR) and thus the quality of MR images during MRgFUS procedures. Acoustic simulations/experiments exhibit transparency of the FUS-Flex coil as high as 97% at 650 kHz. Electromagnetic simulations show a SNR increase of 13x over the body coil. In vivo results show an increase of the SNR over the body coil by a factor of 7.3 with 2x acceleration (equivalent to 11 x without acceleration) in the brain of a healthy volunteer, which agrees well with simulation. These preliminary results show that the use of a FUS-Flex coil in MRgFUS surgery can increase MR image quality, which could yield improved focal precision, real-time intraprocedural anatomical imaging, and real-time 3D thermometry mapping.
更多
查看译文
关键词
Magnetic resonance imaging, coils, ultrasonic transducer arrays