Versatile cryogen-free cryostat for the electromagnetic characterization of superconducting radiofrequency coils

EPJ Techniques and Instrumentation(2020)

引用 2|浏览19
暂无评分
摘要
The use of high temperature superconducting (HTS) radio frequency (RF) coils in Magnetic Resonance Imaging (MRI) greatly improves the signal-to-noise ratio (SNR) in many biomedical applications and particularly in micro-MRI. However, a detailed understanding of the electrical behavior of HTS coils is important in order to optimize their performance through MR experiments. This paper presents a simple and versatile cryogen-free cryostat designed to characterize the RF properties of HTS coils prior to their use in MRI. The cryostat can be used at temperatures from 50 K to 300 K, with a control precision of approximately 3 mK at 70 K, and can measure the RF electrical power transmitted to an HTS coil over a range from 1 μW to 10 W. The quality factor and resonance frequency of the tested HTS coil are determined as a function of the temperature and the power it dissipates. This cryostat also permits the dynamic adjustment of the coil resonance frequency via temperature control. Finally, this study demonstrates that the HTS coil takes less than 12 μs to transit from the superconducting to the dissipative state, which is compatible with MRI requirements.
更多
查看译文
关键词
Superconductivity, HTS coil, Cryostat, High sensitivity